CS/OHN-Analysis (CS/ONH)
XRF Scientific G6 LEONARDO
The G6 LEONARDO is a great choice for process and quality control in primary metal manufacturing and processing, ceramics, and powder metallurgical processes. The three non-metals oxygen (O), nitrogen (N) and hydrogen (H) determine the quality, service life and mechanical properties of all metallic materials.
Benefits
Oxygen, nitrogen and hydrogen analysis by the inert gas fusion method (IGF)
The G6 LEONARDO is based on the inert gas fusion (IGF), which involves fusion of the sample material in a graphite crucible at high temperatures. This principle is also commonly termed gas fusion analysis (GFA) or melt extraction (ME) since the total oxygen, nitrogen and hydrogen is extracted by an inert carrier gas when the sample melts. The aim of an IGF analysis is to determine the sample content of the light elements O, N and H in the trace range with absolute accuracy. That is why G6 LEONARDO applies the Smart Molecule Sequence™ and measures the gases emitted by the sample directly, absolutely unchanged, 1:1, with proven scientific principles for trustworthy results.
Applications
G6 LEONARDO Inert Gas Fusion analysis at work

Iron, Steel, Cobalt, Nickel, Ferroalloys
The G6 LEONARDO combines ease-of-use and robustness to deliver quick and accurate results on oxygen and nitrogen results for quality and process control adjustments. Determining the amount of oxygen and nitrogen is critical for the steel making process – including final alloy adjustment with ferroalloys and production of nickel or cobalt-based alloys.

Copper Production
With a low cost of ownership, the G6 LEONARDO analyzes oxygen with high reliability – even offering precise simultaneous oxygen and hydrogen analysis using the dual element configuration. The level of residual impurities, especially oxygen, is the most important factor when grading the quality of copper. High oxygen levels increase the tendency for hydrogen embrittlement and intergranular failure.

Titanium| Zirconium| Other Refractory Metals
Low density, high strength alloys with excellent corrosion resistance such as titanium – combined with aluminum, vanadium, or molybdenum – are typically used in space, aviation, military, nuclear, biomedical and other demanding applications. The G6 LEONARDO can test the oxygen and nitrogen content as well as hydrogen content to check for hydride precipitation before the subsequent cracking becomes an issue.

Metal Processing
The G6 LEONARDO delivers essential knowledge about the chemical material composition of the metal processing tasks used in aviation, aerospace, automotive, biomedical, electronic and other industries. Whether the metal is stainless steel, aluminum, copper, nickel, titanium, zirconium or special metal alloys, the G6 LEONARDO is up to the task.

Aluminum | Magnesium
With the unique FusionControl and SampleCare, the G6 LEONARDO is the ideal quality control tool for the aluminum industry where oxygen and hydrogen are present in the production and processing of reactive light metal. Oxygen can form insoluble, refractory inclusions while hydrogen dissolved in liquid melt can be trapped during solidification and create hydrogen-filled blisters and voids.

Magnetic Materials
The G6 LEONARDO is an affordable and easy-to-use tool for process control of oxygen – which can reduce magnetic efficiency – and hydrogen which is both necessary to the manufacturing process and potentially problematic. Rare earth metals, such as the sintered Nd-Fe-B magnets show a high tendency to absorb hydrogen which can cause fatal hydrogen embrittlement.

Powder Metallurgy | Additive Manufacturing
The ready-to-use G6 LEONARDO can process high purity materials with strictly defined particle properties such as the metal powders used in additive manufacturing and other powder metallurgical processes where large specific surface leaves them susceptible to oxidation or moisture aging during the production chain.

Glass | Ceramics
The G6 LEONARDO is an excellent tool for the fast analysis of oxygen, nitrogen, and hydrogen in raw materials, in-production products and final products. The unique properties of glass and ceramics are influenced in fundamental ways by the presence of oxygen, nitrogen, or hydrogen during process or in the final product. Whether needed for high performance ceramics in electronics or metal cutting or refractory oxides, nitrides, or carbides the G6 LEONARDO is the way to go.
Specifications
G6 LEONARDO analyzer information
Specification | Benefit | |
Samples | Inorganic, dry, any form or shape | Flexibility |
Sample Dimensions | 8 x 8 mm cross section (pieces, chips, pins, granules), different sample ports available for special applications | Wide variety of sample materials |
Furnace | Water-cooled Electrode Furnace with FusionControl | Solid high temperature operation for refractories, precise temperature control |
Detectors Oxygen Hydrogen | Nondispersive IR absorption of CO, no moving parts, multiple range detectors with reference beam and on-board linearization Thermal conductivity detector with reference channel and | Reliable, maintenance-free performance without drift Reliable, adjustable ranges, drift free |
Cooling Water | 4 l/min at 3 bar supply pressure | Fits on every usual tap water line, tap, water-saving |
SampleCare™ EZ Drive | Lower electrode movement by reliable electronic motion module | Robust, cost-saving operation without working gas. Automatic, torque limited readjustment ensures ideal crucible:electrode contact and safety for crucible, furnace and operator. |
Sample Port | Rotating sample port integrated into water cooled furnace block | Shields sensitive samples from heat input and hydrogen losses before analysis. Allows capsule-free measurement of grains without jamming. |
Particle Filtering | High-capacity dust trap followed by in-line particle filtering | Stable analytical results with reduced maintenance and |
Carrier Gas | Ar (O, N), He (O, N), N2 (H) each 99.995 % purity | Flexible carrier gas choice without compromise |
Noise Emission | < 55 dB(A) peak | Enjoy the silence |
Calibration | Single and multipoint calibration, advanced weighting algorithm with error bars and range merging | Easy simultaneous calibration of all ranges with |
Reagents | - Magnesium perchlorate | |
Power Supply | 400 - 480 VAC (± 10 %), 50-60 Hz, 32 A, 7 kVA, 1 Phase plus N or 2 Phase, grounding wire | Compatible to worldwide power and current |
Models Available | Single or dual element O, N, H, ON, OH | Ideal tool for process and quality control |
Dimensions and Weight | 71 x 66 x 78 cm (W x D x H), ~145 kg | Compact form, easy to maintain |
*) Depends on configuration